Type J, K or T Thermocouple Thermometer from Barnant
Hold button freezes display for later reading
Easy to use, field calibration "capture" mode for improved system accuracy
Accuracy: ±0.9°F (±0.5°C) above -99.9 ±2.0°F (1.0°C) below -99.9 plus ± 0.2% of reading
Selectable °C or °F scale
UL & cUL rated Intrinsically safe-Certified for Class I, Division 1; Groups A, B, C, and D hazardous (classified) locations
Sealed Silicone rubber keypad
Rated IP 54 for splash & dust resistance
ABS/polcarbonate impact resistant case
Fold out stand for bench top operation
Meets CE requirements for electromagnetic compatibility
Bright, easy to read LCD
3 year warranty
Manufactured in U.S.A.
Easy-to-use, minimally featured Thermocouple Thermometer is ideal for your basic temperature monitoring applications. A Thermocouple Thermometer is available for either J, K, or T type thermocouples. Four button keypad: On/Off, °C/°F for °C or °F scale, Hold and CAL. CAL button - easy one button, "capture" mode, for a single (ice) point field calibration offset. Field Calibration lockout feature prevents tampering and protects factory settings.
K Thermocouples are the most popular and have the widest temperture range, J Thermocouples offer ruggedness and reliability, T Thermocouples have the narrowest range but are ideal for cryogenics, food service and research applications under 200°F.
Type J, K or T Thermocouple Thermometer Specifications
ALL THERMOMETERS:
Power: two AA batteries (included)
Battery life: minimum 500 hours
Low battery indication: yes
Dimensions: 3.3"W x 6.2"H x 1.2"D (84 mm W x 158 mm H x 30 mm D)
Shipping Weight: 0.7 lbs (0.4kg)
Warranty: 3 years
Housing: ABS/polcarbonate impact resistant case with sealed silicone rubber keypad and built-in tilt stand Ratings: UL & cUL rated Intrinsically safe - Certified safe for Class I, Division 1; Groups A, B, C, and D hazardous (classified) locations. Meets CE requirements for electromagnetic compatibility
|
Cat. no.
|
Type
|
Range
|
Accuracy
|
Resolution
|
Functions
|
Display
|
600-1000
|
J
|
-328 to 1832°F (-200 to 1000°C)
|
above -99°: ±(0.2% of reading and ±0.9°F (± 0.5°C)) below -99° ±(0.2% of reading and ±2°F (±1°C))
|
0.1° between -99.9° and 299.9°: 1° below -99.9°
|
On/off, hold calibration, and °C/°F
|
3 3/4 digit LCD, 1/2"H
|
600-1010
|
K
|
-418 to 2501°F (-250 to 1372°C)
|
600-1020
|
T
|
-418 to 752°F (-250 to 400°C)
|